Learn from the record. Every FDA recall is a quality signal.
Live drug, device & food recall data from the FDA (FDA enforcement) — food spanning dietary supplements, cosmetics & animal food — mapped to the GxP quality domains SPEQ assesses, so you can see where control breaks down across every regulated industry.
Keep FDA on your watchlist.
Drug, device and food enforcement actions and recall classifications, as FDA posts them. Track FDA and SPEQ sends you one weekly email with what it published — nothing on a quiet week — so the bodies you answer to sit in one place instead of across a dozen tabs.
Free, and no card. A free account holds up to 25 tracked records, plus saved items and notes. Turn the weekly email off any time from your saved items.
Domain mapping is a SPEQ interpretation of the FDA recall reason, not an official FDA classification.
Presence of Foreign Tablets/Capsules
Failed Impurities/Degradation Specifications.
Lack of assurance of sterility.
Lack of process control
Presence of Foreign Tablets/Capsules
Products may be contaminated with Listeria monocytogenes.
Undeclared colors (FD&C Blue No. 1 and FD&C Yellow No. 5)
Undeclared soy.
defective packaging; packaging can rupture/explode during cooking.
Undeclared Lobster
Foreign object: May contain pieces of metal
Foreign object: May contain pieces of metal
May contain undeclared aspartame.
Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.
Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.
Delivery Catheter may have sleeve liner separation that results in difficulty or inability to deploy heart valve device.
Delivery Catheter may have sleeve liner separation that results in difficulty or inability to deploy heart valve device.
Delivery Catheter may have sleeve liner separation that results in difficulty or inability to deploy heart valve device.
Medtronic identified through donor record review that the donor for Medtronic batch A62881 did not meet all Medtronic donor eligibility requirements related to the presence of microbial organisms. While the batch passed finished product sterility testing Medtronic is retrieving any unused product manufactured from the donor tissue. Donor bone tissue tested positive for Clostridium butyricum and while tissue did undergo gamma radiation, if implanted, there is a possibility of clostridium infection
Medtronic identified through donor record review that the donor for Medtronic batch A62881 did not meet all Medtronic donor eligibility requirements related to the presence of microbial organisms. While the batch passed finished product sterility testing Medtronic is retrieving any unused product manufactured from the donor tissue. Donor bone tissue tested positive for Clostridium butyricum and while tissue did undergo gamma radiation, if implanted, there is a possibility of clostridium infection.
Potential for quality control failures with Klebsiella pneumoniae ATCC BAA-1705 when using the CPO Ambler classification feature.
Potential for quality control failures with Klebsiella pneumoniae ATCC BAA-1705 when using the CPO Ambler classification feature.
Stimulator device contains a defective printed circuit board assembly (PCBA) that may cause the device to unintentionally deliver electrical stimulation to the patient, or allow an unintended electrical current to reach the patient.
Cables are not in conformance with the product specifications required in the production of the HP Coil Assembly.
Device was incorrectly manufactured with a guidewire exit port (referred to in the IFU as the RX port), which is not a design feature associated with this device configuration.
Affected kits were assembled with pediatric nasal cannula instead of the intended adult nasal cannula.
Firmware issue may result in intermittent loss or degradation of powered functionality when paired with compatible stretcher. The issue may lead to the system not operating in powered mode, or exhibiting reduced powered performance. Issue may result in interruption or delay in patient transport.
The pelvic fixation screws failed to meet acceptance criteria for tube packaging seal integrity testing; therefore, sterility assurance cannot be ensured.
Subrecall of a kitted device from SafeGuard Medical due to internal testing identified an issue affecting the device's sterile pad, potentially causing the sterile pad to degrade over time. The degradation may cause remnants of the nonwoven pad material to dislodge on the wound or in close proximity to the wound.
Due to potential contaiminated of spefic lots of pneumonia and joint infection panels that could result in false positive results
Due to potential contaiminated of spefic lots of pneumonia and joint infection panels that could result in false positive results
Due to potential contaiminated of spefic lots of pneumonia and joint infection panels that could result in false positive results
Potential issue involving motor shaft breakage combined with reduced performance of the Single Fault Safety (SFS) system may lead to instances where the lift strap lowers more quickly than expected during use. Patients are at some risk for injury if a significant fall/drop were to occur due to this issue.
Silk sutures, indicated for use in cardiovascular, ophthalmic and neurological procedures, packaged with alcohol-based fluid, not part of product design, that if discovered prior to use may cause surgical delay. If used, fluid could cause reduced suture tensile strength or knot security, tissue irritation/inflammatory tissue response, and treatment failure, including wound dehiscence/bleeding.
If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.
If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.
If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.
Due to labeling error, bone screws specified as fully threaded, are actually partially threaded screws.
The potential for the devices to be mislabeled with the incorrect length.
Vial-to-vial variability in MAS¿ Omni"IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, Lot OPRO29061A results in low Parathyroid Hormone (PTH) recovery, which may lead to a delay in reporting patient PTH results.
The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual preventive maintenance (PM) requirements without adequate reliability evaluation. When the check valve is not replaced annually, it may allow lamina fluid to migrate internally and leak onto the table or floor, potentially resulting in exposure to physical hazards (slip/fall), chemical hazards (lamina fluid exposure), biohazardous exposure, property damage, and delayed patient results.
The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual preventive maintenance (PM) requirements without adequate reliability evaluation. When the check valve is not replaced annually, it may allow lamina fluid to migrate internally and leak onto the table or floor, potentially resulting in exposure to physical hazards (slip/fall), chemical hazards (lamina fluid exposure), biohazardous exposure, property damage, and delayed patient results.
In kit sub-lot 2530M-01 the date of expiration printed on the outer labeling may indicate the wrong expiration date of July 22, 2026 while the correct expiration date is July 13, 2026. If used the system will display an error message and prevent the operator from running the test. Resulting in delayed results.
The female Luer lock adapter component in the affected device lots may not have been correctly seated due to misalignment and inadequate torquing during manufacturing. Potential events that may occur as a result of this issue include increased procedural time and inadequate treatment.
Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.
Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.
Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.
Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.
Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.
Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.
Following a Chromium-based web-browser, report and displayed images may not match.
The implantable clip may experience difficulties deploying, which could prevent the reflector from functioning as intended during use.
Labeling not compliant with 21 CFR 801 and steam sterilization IFU has been updated.
Discrepancy in the shelf-life of a component
In vitro diagnostic analyzer had a risk of potentially inaccurate results requiring changes to instructions for use and: 1) Sample types, including whole blood samples no longer suitable for testing FHbF and ctBil, 2) Reportable ranges for some parameters, 3) Analyzer configuration, and 4) limitations and warnings.
Lack of Assurance of Sterility
Lack of Assurance of Sterility
Lack of Assurance of Sterility
Lack of Assurance of Sterility
Lack of Assurance of Sterility
Lack of Assurance of Sterility
Lack of Assurance of Sterility
Lack of Assurance of Sterility
Failed stability specifications
Failed Content Uniformity Specifications
Failed Impurities/Degradation Specifications
Presence of particulate matter.
Lack of Assurance of Sterility: Related to formulation using bacteriostic water; Related to processing equipment used in compounded product.
Lack of Assurance of Sterility: Related to formulation using bacteriostic water; Related to processing equipment used in compounded product.
Lack of Assurance of Sterility: Related to formulation using bacteriostic water; Related to processing equipment used in compounded product.
Lack of Assurance of Sterility: Related to formulation using bacteriostic water; Related to processing equipment used in compounded product.
Lack of Assurance of Sterility: Related to processing equipment used in compounded product.
Lack of Assurance of Sterility: Related to processing equipment used in compounded product.
Lack of Assurance of Sterility: Related to processing equipment used in compounded product.
Lack of Assurance of Sterility: Related to processing equipment used in compounded product.
Lack of Assurance of Sterility: Related to processing equipment used in compounded product.
Lack of Assurance of Sterility: Related to formulation using bacteriostic water; Related to processing equipment used in compounded product.
Lack of Assurance of Sterility: Related to formulation using bacteriostic water; Related to processing equipment used in compounded product.
Lack of Assurance of Sterility: Related to formulation using bacteriostic water; Related to processing equipment used in compounded product.
Lack of Assurance of Sterility: Related to formulation using bacteriostic water; Related to processing equipment used in compounded product.
Lack of Assurance of Sterility: Related to formulation using bacteriostic water; Related to processing equipment used in compounded product.
Lack of Assurance of Sterility: Related to formulation using bacteriostic water; Related to processing equipment used in compounded product.
Lack of Assurance of Sterility: Related to formulation using bacteriostic water; Related to processing equipment used in compounded product.
Failed Impurities/Degradation Specifications
Failed Impurities/Degradation Specifications
Failed Impurities/Degradation Specifications
Lack of Assurance of Sterility
Lack of Assurance of Sterility
Seeds may have been contaminated with E. coli
Seeds may have been contaminated with E.coli
Potential contamination with foreign objects: mice.
Bean Dip Trays contains recalled Peperoncini (pepper) which was recalled due to pest issue.
Undeclared Yellow No. 6 and/or Yellow No. 5
Undeclared Milk.
The chicken salad sandwich has an improperly declared allergen (egg). While egg is included in the ingredients list on the package label, it is not included in the allergen Contains statement.
Product contains alfalfa seed recalled by the supplier due to potential E. coli and/or Salmonella contamination.
Contamination with Listeria monocytogenes
Undeclared milk, cheese declared
Product contains undeclared Sildenafil and Tadalafil.
Product contains undeclared Sildenafil and Tadalafil.
Staphylococcus aureus detected in product.
Undeclared soy allergen: soybean oil and soy lecithin
Undeclared soy allergen: soybean oil and soy lecithin
Undeclared soy allergen: soybean oil and soy lecithin
Undeclared soy allergen: soybean oil and soy lecithin
Product contains undeclared soybean oil and soy lecithin in the Ingredient Statement. Soy was not declared in the Contains statement.
Soy was not declared in the Contains statement.
Egg was not declared in the Contains statement.
Potential contamination due to E. coli O26:H11.
An internal data-entry error caused kit lot 333606 to be labeled with an expiration date of March 31, 2027, instead of February 28, 2027.
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
Due to a manufacturing issue (inadequate weld) for patient head positioning system.
Laparoscopy packs labeled as "latex free" were found to contain a latex urethral catheter
Software issues may affect the safety and performance of Spectral CT system(s). The issues may result in image artifacts, wrong slice thickness, or unintended system motion.
Kits contain Microstream components. Medtronic issued an update to the Instructions for Use of these components due reports of difficulty in disconnecting the adapter from the patient's endotracheal tube to perform a procedure. The issue can result in a delay to treatment and unintended extubation. The potential exists for respiratory failure, hypoxia, aspiration, or low oxygen saturation.
Lack of 510(k) clearance.
Compression bandage may be degraded, which may cause unwanted foreign material to be introduced inside the wound, which may lead to wound exploration and irrigation, formation of scar tissue, and infection.
Potential for false negative patient results.
Beckman Coulter has identified a software issue affecting DxC 500i Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system default, a subsequent configuration change and activation of the draft configuration in the System Configuration page may cause the analyzer to revert the test units to the system default setting without notifying the user. The system does Pre-Review Reminders Include OII and CDRHRecallGroup@fda.hhs.gov in any discussions on change/elevation of recalls to Class I classification and situations involving additional public communications (FDA and/or firm issued). Post-Review Reminders Final Check: Ensure your documentation accurately reflects the final recommendation prior to signature. Complete CTS: Recall Product tab, Additional Recall Event Info, Regulation, and Act. Upload PDFs of Consult, Review Memo, and Classification Letter for Class I Recall in DocMan. [Click to enter extra reminder]To add reminders, type the reminder in the text field to the left then press the Tab key. Doc ID 03534.05.06 2 of 8 not provide a warning or notification to alert the user that the configured units have reverted to the default units. This issue can result in falsely elevated reported results.
Beckman Coulter has identified a software issue affecting DxC 500 AU Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system default, a subsequent configuration change and activation of the draft configuration in the System Configuration page may cause the analyzer to revert the test units to the system default setting without notifying the user. The system does Pre-Review Reminders Include OII and CDRHRecallGroup@fda.hhs.gov in any discussions on change/elevation of recalls to Class I classification and situations involving additional public communications (FDA and/or firm issued). Post-Review Reminders Final Check: Ensure your documentation accurately reflects the final recommendation prior to signature. Complete CTS: Recall Product tab, Additional Recall Event Info, Regulation, and Act. Upload PDFs of Consult, Review Memo, and Classification Letter for Class I Recall in DocMan. [Click to enter extra reminder]To add reminders, type the reminder in the text field to the left then press the Tab key. Doc ID 03534.05.06 2 of 8 not provide a warning or notification to alert the user that the configured units have reverted to the default units. This issue can result in falsely elevated reported results.
Diagnostic electrophysiology (EP) catheter may have a breached sterile barrier on the straight seal side of the primary pouch, which if used could lead to particulate embolization, infection, thrombus or ganuloma formation from particulate matter.
The front fork stem bolt/wheel assembly may disconnect from the rollator frame tubing. If the front fork stem bolt/wheel assembly disconnects from the rollator frame tubing, the front wheels may become loose or detach from the rollator. If the front wheels become loose or detach during use, there is risk of severe adverse events. The immediate consequences could include falls resulting in head trauma, bone fractures, or other critical injuries.
The front fork stem bolt/wheel assembly may disconnect from the rollator frame tubing. If the front fork stem bolt/wheel assembly disconnects from the rollator frame tubing, the front wheels may become loose or detach from the rollator. If the front wheels become loose or detach during use, there is risk of severe adverse events. The immediate consequences could include falls resulting in head trauma, bone fractures, or other critical injuries.
Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed verification test results in one or more of 9 functional tests. The unverified/failed pumps released to users potentially have one or more of the following issues: inoperable pump at setup, lack of protection for moisture exposure, slow battery charging, disconnection issues with other system components (i.e., continuous glucose monitoring sensor or MiniMed mobile app), reduced ability to detect occlusion events, compromised audio for pump alarms/alerts, and/or inaccurate insulin delivery.
Due to a cybersecurity issue associate with a NULL Pointer Dereferences vulnerabiity. If exploited would potentially cause disruption of the centralized monitoring of multiple beds and the visualization of alarms.
Personalized pelvic implants inadvertently mixed up and distributed to incorrect customers.
Personalized pelvic implants inadvertently mixed up and distributed to incorrect customers.
Product may be potentially contaminated with a lubricant used during the manufacturing process.
Product may be potentially contaminated with a lubricant used during the manufacturing process.
Product may be potentially contaminated with a lubricant used during the manufacturing process.
Product may be potentially contaminated with a lubricant used during the manufacturing process.
Product may be potentially contaminated with a lubricant used during the manufacturing process.
Product may be potentially contaminated with a lubricant used during the manufacturing process.
Product may be potentially contaminated with a lubricant used during the manufacturing process.
Product may be potentially contaminated with a lubricant used during the manufacturing process.
Product may be potentially contaminated with a lubricant used during the manufacturing process.
Product may be potentially contaminated with a lubricant used during the manufacturing process.
Product may be potentially contaminated with a lubricant used during the manufacturing process.
Product may be potentially contaminated with a lubricant used during the manufacturing process.
Product may be potentially contaminated with a lubricant used during the manufacturing process.
Product may be potentially contaminated with a lubricant used during the manufacturing process.
Product may be potentially contaminated with a lubricant used during the manufacturing process.
During use of an affected tube, the operator may encounter increased resistance when inserting the pipette. The resulting pressure buildup may cause the tube cap to detach or be forcibly ejected, potentially resulting in the release of blood and exposure of the operator or surrounding work area to a biological hazard (blood).
Loosening of the bolts securing the longitudinal movement mechanism may cause the tabletop to shift, resulting in [positional/imaging/patient] misalignment and a longitudinal motion error
Certain monitor operator's manuals contain descriptions that may be insufficient or may require clarification regarding selected physiological alarm functions
Certain monitor operator's manuals contain descriptions that may be insufficient or may require clarification regarding selected physiological alarm functions
Certain monitor operator's manuals contain descriptions that may be insufficient or may require clarification regarding selected physiological alarm functions
Certain monitor operator's manuals contain descriptions that may be insufficient or may require clarification regarding selected physiological alarm functions
Certain monitor operator's manuals contain descriptions that may be insufficient or may require clarification regarding selected physiological alarm functions
Certain monitor operator's manuals contain descriptions that may be insufficient or may require clarification regarding selected physiological alarm functions
Certain monitor operator's manuals contain descriptions that may be insufficient or may require clarification regarding selected physiological alarm functions
Certain monitor operator's manuals contain descriptions that may be insufficient or may require clarification regarding selected physiological alarm functions
Devices failed in process testing related to the amount of pull force it can sustain prior to the balloon components failing.
A software issue may result in unintended mismatch of patient information on manually printed images. Incorrect patient information appearing on printed images could create a documentation discrepancy or incorrect association of clinical images with a patient record.
A navigation system used to aid in selection and positioning of orthopedic implant system components has a workstation that may contain a 15-inch MacBook Pro battery, subject to an Apple battery recall, that may overheat and pose a fire safety risk.
Failed Dissolution Specifications: below specification
Defective Container; packets were found to be either empty or partially full.
Lack of Assurance of Sterility
Temperature Abuse: Product sat in a non-temperature-controlled FedEx trailer at ambient outdoor conditions for 25.5 hours.
Crystallization
Failed dissolution specifications.
Microbial contamination of topical product.
Microbial contamination of topical product.
Discoloration: consumer complaints received for discolored tablets
Undeclared Yellow 5 & 6, Red 3, and Blue 1
Product may contain undeclared sugar.
This recall was initiated due to potential Salmonella contamination in diced jalape¿no received from the supplier
Product contains garlic essential oil not approved for culinary use
Undeclared allergen ingredient (fish)
Undeclared shellfish (crab shell): source of ingredient Glucosamine sulfate
Product contains elevated levels of lead.
Potential Salmonella contamination.
Products contain alfalfa seed recalled by the supplier and may be contaminated with E. coli and/or Salmonella (downstream recall)
Products contain alfalfa seed recalled by the supplier and may be contaminated with E. coli and/or Salmonella (downstream recall)
Products contain alfalfa seed recalled by the supplier and may be contaminated with E. coli and/or Salmonella (downstream recall)
Products contain alfalfa seed recalled by the supplier and may be contaminated with E. coli and/or Salmonella (downstream recall)
Products contain alfalfa seed recalled by the supplier and may be contaminated with E. coli and/or Salmonella (downstream recall)
Products contain alfalfa seed recalled by the supplier and may be contaminated with E. coli and/or Salmonella (downstream recall)
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Due to manufacturing issues with the liquid skin adhesive containing in surgical convenience trays/kits.
Retrospectively reported correction for an instrument intended for targeted sequencing of DNA libraries from human genomic DNA used with in vitro diagnostic assays has a server with a security vulnerability.
The Bolus tracking scan application may crash with a system error message. Bolus Tracker shots may continue to occur even if the contrast threshold is reached. This may result in the patient receiving additional radiation.
Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire. Additional products added to 11/26/2025 recall.
Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire. Additional products added to 11/26/2025 recall.
Potential cracks in the hose, which could lead to leakage or the hose falling off.
Products were distributed without the required Maintenance Supplement, which describes an extra check every 3 months in addition to the monthly check.
hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.
hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.
hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.
hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.
Breast biopsy and tissue removal system may experience the following errors: 1) Probe calibration and /or handheld driver failure (code HH01-00020009) , 2) Pressure sensor (code E01-0054000E), which prompts the user to restart and re-calibrate; errors may lead to interruption or prolongation of the biopsy procedure, or the need for additional tissue sampling or repeat biopsy.
Software system may associate image data from one patient with another patient's record, potentially resulting in incorrect diagnosis or treatment.
Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
Positron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
Gloves may exhibit defects such as tears, missing fingers, and debris. Use of defective gloves may lead to healthcare personnel and patient exposure to infectious pathogens leading to infection or cross-contamination, or exposure to chemical agents leading to irritation or systemic toxicities.
Gloves may exhibit defects such as tears, missing fingers, and debris. Use of defective gloves may lead to healthcare personnel and patient exposure to infectious pathogens leading to infection or cross-contamination, or exposure to chemical agents leading to irritation or systemic toxicities.
Gloves may exhibit defects such as tears, missing fingers, and debris. Use of defective gloves may lead to healthcare personnel and patient exposure to infectious pathogens leading to infection or cross-contamination, or exposure to chemical agents leading to irritation or systemic toxicities.
Kits contain manufacturer-recalled perineal care (pericare) wipes within urinary catheter trays, which may be contaminated with a lubricant used during the manufacturing process. Use of a contaminated wipe may lead to localized skin effects such as irritation, erythema, or mild inflammation. Failure to perform pericare due to disposal of contaminated wipes may increase risk of localized infection or catheter-associated urinary tract infection (CAUTI).
Kits contain manufacturer-recalled perineal care (pericare) wipes within urinary catheter trays, which may be contaminated with a lubricant used during the manufacturing process. Use of a contaminated wipe may lead to localized skin effects such as irritation, erythema, or mild inflammation. Failure to perform pericare due to disposal of contaminated wipes may increase risk of localized infection or catheter-associated urinary tract infection (CAUTI).
Kits contain manufacturer-recalled perineal care (pericare) wipes within urinary catheter trays, which may be contaminated with a lubricant used during the manufacturing process. Use of a contaminated wipe may lead to localized skin effects such as irritation, erythema, or mild inflammation. Failure to perform pericare due to disposal of contaminated wipes may increase risk of localized infection or catheter-associated urinary tract infection (CAUTI).
The kits contain tube sets with handpieces that may leak during use. The impacted devices can have the following failures: fluid leakage from the hand piece that can pool on the floor. The impact can range from user dissatisfaction to the potential for the user of slip and fall.
The kits contain tube sets with handpieces that may leak during use. The impacted devices can have the following failures: fluid leakage from the hand piece that can pool on the floor. The impact can range from user dissatisfaction to the potential for the user of slip and fall.
The kits contain tube sets with handpieces that may leak during use. The impacted devices can have the following failures: fluid leakage from the hand piece that can pool on the floor. The impact can range from user dissatisfaction to the potential for the user of slip and fall.
The kits contain tube sets with handpieces that may leak during use. The impacted devices can have the following failures: fluid leakage from the hand piece that can pool on the floor. The impact can range from user dissatisfaction to the potential for the user of slip and fall.
There are identified discrepancies between the sterilization parameters listed in the Instructions for Use and the sterilization parameters listed in the Japanese package insert.
There are identified discrepancies between the sterilization parameters listed in the Instructions for Use and the sterilization parameters listed in the Japanese package insert.
Data: U.S. FDA. Quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Recall details are as reported by the FDA.
That is what FDA published. A free account files it on your watchlist and sends one weekly email with whatever lands next.
Create a free account