Due to a firmware defect, the portable organ transfer device datalogger may lose its configuration, resulting in a blank screen and loss of displayed preservation parameters.
Paragonix Technologies Inc. KidneyVault Portable Renal Perfusion System REF# PRGNX-8008-001
All Serial Numbers/UDI: 00850005470342 Software Version: Paragonix App Ver. 6.1.1/6.1.2