Fort Defiance Industries, LLC
What this record is not
A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.
The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.
Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131
Lot Code: All serial numbers/ GTIN: (01) 00862992000327