← FDA ENFORCEMENT
Z-3034-2026Class II

IMRIS Imaging Inc

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
IMRIS Imaging Inc
Product type
Device
Status
Ongoing
Report date
2026-09-02
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, China, Japan, Sweden.
Firm state
MN
Product quantity
13 units
Initiated by
Voluntary: Firm initiated
Reason for recall

The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.

Product description

ORT300, operating room table, ORT300 Modular Table Main Unit Assy Part Numbers 114093-000 and 114093-600

Code / lot info

114093-000 - UDI-DI: 00857534006356, Serial Numbers: 10005695, 10002998, 10002997, 10005456, 10005612, 10004372, 10005613, 10002841, 10011119; 114093-600 - UDI-DI: 00857534006462, Serial Numbers: 10002996, 10004515, 10007411, 10007376