← FDA ENFORCEMENT
Z-2904-2026Class II

TISSIUM SA

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
TISSIUM SA
Product type
Device
Status
Ongoing
Report date
2026-08-12
Distribution
US Nationwide distribution in the states of CA, FL, IL, NJ, OH, PA.
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Validation & Qualification
Reason for recall

Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.

Product description

Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2075 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No

Code / lot info

Lot Code: Model No NCOCO2075; UDI-DI ¿03760279380100; Lot Number 250700016, 251000048, 251000049, and 260300043