← FDA ENFORCEMENT
Z-2916-2026Class II

Spacelabs Healthcare, Inc.

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Spacelabs Healthcare, Inc.
Product type
Device
Status
Ongoing
Report date
2026-08-19
Distribution
Worldwide - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WY and the countries of CANADA, FRANCE, ITALY, MALAYSIA, THAILAND.
Firm state
WA
Product quantity
6074
Initiated by
Voluntary: Firm initiated
Reason for recall

Due to off-the-shelf display port component that has audio/video failures

Product description

Brand Name: Xhibit Product Name: Xhibit Central Station Model/Catalog Number: 96102 Software Version: N/A Product Description: The Spacelabs Healthcare (Spacelabs) Xhibit Central Station, Model 96102. Xhibit offers remote surveillance of patient data for those patients connected to a Spacelabs Healthcare bedside monitors and telemetry systems. The DisplayPort audio/video splitter allows users to extend the audio/video source from the Xhibit to additional external displays or monitors. This extension of audio/video is for viewing only and does not allow interaction with the Xhibit. Component: The DisplayPort audio/video splitter is a 3rd party external accessory to Xhibit.

Code / lot info

Model/Catalog Number: 96102 DisplayPort Splitter Part Number: 010-1975-00 UDI-DI codes: 1084152212706 10841522100345 Serial Numbers: Pending

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →