Z-2950-2026Class II

Covidien, LLC

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Covidien, LLC
Product type
Device
Status
Ongoing
Report date
2026-08-26
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, and WV. The countries of Argentina, Australia, Austria, Belgium, Canada, Colombia, Costa Rica, Denmark, France, Germany, Hong Kong, Ireland, Israel, Italy, Netherlands, New Zealand, Northern Ireland, Norway, Portugal, South Korea, Spain, Sweden, Switzerland, United Kingdom.
Firm state
CA
Product quantity
327
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Validation & Qualification
Reason for recall

The system's 24-hour usage limit results in catheters incorrectly flagged as expired, making them unavailable for clinical procedures.

Product description

EndoFLIP System that includes: ENDOFLIP 1 SYSTEM EF-100 05391530810050 10884521809307 REFURBISHED SYSTEM RFG-EF100 ENDOFLIP 1 10884521808096 05391530810166 DL ENDOFLIP 1 10884521786684 Endoscopic measuring system, including catheter components.

Code / lot info

ENDOFLIP 1 SYSTEM EF-100 05391530810050 10884521809307 REFURBISHED SYSTEM RFG-EF100 ENDOFLIP 1 10884521808096 05391530810166 DL ENDOFLIP 1 10884521786684 Catheters used after January 1, 2026.