← FDA ENFORCEMENT
Z-2974-2026Class II

Beckman Coulter Inc.

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Beckman Coulter Inc.
Product type
Device
Status
Ongoing
Report date
2026-08-26
Distribution
Worldwide distribution - US Nationwide.
Firm state
CA
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Data Integrity
Reason for recall

a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.

Product description

Beckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE, Part Number C63519; 2) AU w/ISE, Part Number C63520; DxC 500 AU Clinical Chemistry Analyzers

Code / lot info

1) Part Number C63519, UDI/DI 14987666545058 2) Part Number C63520, UDI/DI 14987666545065