← FDA ENFORCEMENT
Z-3033-2026Class II

IMRIS Imaging Inc

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
IMRIS Imaging Inc
Product type
Device
Status
Ongoing
Report date
2026-09-02
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, China, Japan, Sweden.
Firm state
MN
Product quantity
6 units
Initiated by
Voluntary: Firm initiated
Reason for recall

The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.

Product description

ORT200, operating room table, including the ORT200 Covered Table Assembly, Reconditioned, Part Number 113821-600 and the ORT200 50/60Hz Covered Table Assy V2 Part Number 114148-000

Code / lot info

113821-600 - Serial Numbers: 10003754, 10006997; 114148-000 - UDI-DI: 00857534006349, Serial Numbers: 10005189, 10005190, 10004463, 10004997

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →