← FDA ENFORCEMENT
Z-2880-2026Class III

Siemens Healthcare Diagnostics, Inc.

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product type
Device
Status
Ongoing
Report date
2026-08-12
Distribution
Worldwide - US Nationwide distribution in the states of CA and VA. The country of Japan.
Firm state
DE
Product quantity
8 US; 44 OUS
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Validation & Qualification
Reason for recall

Out-of-specification sodium azide concentration in Syva Emit Disopyramide Enzyme Immunoassay Calibrator Lot U1. While the assay results were not impacted, the overall health risk to lab technicians due to this issue is low, as exposure is limited by small volumes and a 0.2% post-reconstitution sodium azide concentration, with the worst outcome being mild skin or eye irritation and no serious harm anticipated.

Product description

Syva Emit Disopyramide Assay Enzyme Immunoassay, Lidocaine SMN 10445370 Catalog No 6N319UL

Code / lot info

UDI-DI 00842768001406 Lot U1