← FDA ENFORCEMENT
Z-3257-2026Class II

Baxter Healthcare Corporation

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Baxter Healthcare Corporation
Product type
Device
Status
Ongoing
Report date
2026-09-30
Distribution
US Nationwide distribution in the states of Colorado and Florida.
Firm state
IL
Product quantity
2
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Environmental Monitoring
Reason for recall

Contamination within the release test equipment flow path could have affected the accuracy of the nebulizer flow rate. Affected devices returned to firm and testing confirmed devices meet specifications.

Product description

Brand Name: VOLARA System Product Name: VOLARA, HC, NA A Model/Catalog Number: M08594A Software Version: N/A Product Description: VOLARA, HOMECARE, NORTH AMERICA Component: No

Code / lot info

Lot Code: UDI: 00887761981690 Serial Numbers: W078SY3553 and Z309SY7853