Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
Triathlon Hinge Femoral Prep Tray. Catalog Number: 6543-6-514.
Catalog Number: 6543-6-514. GTIN: 07613327562859. All lot numbers used prior to the correction date of May 8, 2026 are impacted.