← FDA ENFORCEMENT
Z-3215-2026Class II

Heidelberg Engineering GmbH

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Heidelberg Engineering GmbH
Product type
Device
Status
Ongoing
Report date
2026-09-30
Distribution
Worldwide - US Nationwide distribution in the state of Massachusetts and the countries of Australia, Costa Rica, Germany, France, India, Israel, Italy, Canada, Mexico, Netherlands, Austria, Philippines, Poland, Saudi Arabia, Switzerland, Spain. South Africa. Thailand, Czech Republic, United Kingdom, Vietnam.
Product quantity
87
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Validation & Qualification
Reason for recall

Wheels may come loose when module is being pushed or pulled during movement or transportation, affecting stability of the stand. The stand may tip over as a result. Falling stand may cause impact or crushing injuries, particularly to the feet, legs, hands or arms of the operator or nearby persons.

Product description

Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex Module Model/Catalog Number: S3610, S2610, S2000, S1610 multimodal imaging platform optimized for the posterior segment of the eye

Code / lot info

UDI-DI 04059237000235 Lot Code: Spec-07265, Spec-11121, Spec-11281, Spec-12096, Spec-13305, Spec-14403, Spec-15282, Spec-15587, Spec-20071, Spec-20348, Spec-20351, Spec-20591, Spec-20605, Spec-23473, Spec-23941, Spec-23950, Spec-23975, Spec-24507, Spec-24558, Spec-24582, Spec-24611

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →