Microgenics Corporation
What this record is not
A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.
Vial-to-vial variability in MAS¿ Omni"IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, Lot OPRO29061A results in low Parathyroid Hormone (PTH) recovery, which may lead to a delay in reporting patient PTH results.
MAS Omni IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, REF: OPRO-101; MAS Omni IMMUNETM PRO Liquid Assayed Integrated Immunoassay Control, Sample Pack, REF: OPRO-SP
REF/UDI-DI/Lot(Expiration): OPRO-101/00884883005327/OPRO29061A(2029-06-30); OPRO-SP/00884883005358/OPRO2906SA(2029-06-30)
Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →