Medline Industries, LP
What this record is not
A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.
Kits may exhibit open seals in sterile pouch packaging. Open seal may compromise product sterility. If not detected prior to use, issue may lead to contamination of sterile field or patient infection if product is used.
Convenience kits MINOR PACK DYNJ58421A SPRINGHILL MINOR PACK DYNJ40734F
DYNJ58421A UDI-DI 10193489970487 Lot 25HBQ049 DYNJ40734F UDI-DI 10198459162084 Lots 25GBO299 25IBS607
Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →