← FDA ENFORCEMENT
Z-3127-2026Class I

Medline Industries, LP

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Medline Industries, LP
Product type
Device
Status
Ongoing
Report date
2026-09-23
Distribution
US Nationwide distribution.
Firm state
IL
Product quantity
47848 kits
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Environmental Monitoring
Reason for recall

The kits contain certain lots of sterile applicators which may contain fungal contamination that may occur under certain environmental conditions, allowing the growth of Aspergillus penicillioides. Contamination of skin preparation products with Aspergillus penicillioides may lead to serious adverse health consequences, including systemic infection, sepsis, illness, and death.

Product description

Medline Convenience Kits: 1) DYNDB3086D, Catalog Number: DYNDB3086D; 2) DYNDB3100, Catalog Number: DYNDB3100; 3) DYNDH1619A, Catalog Number: DYNDH1619A; 4) DYNDH1620A, Catalog Number: DYNDH1620A; 5) DYNDH1694A, Catalog Number: DYNDH1694A; 6) DYNDH1695, Catalog Number: DYNDH1695; 7) DYNDH1708A, Catalog Number: DYNDH1708A; 8) DYNDH1714A, Catalog Number: DYNDH1714A; 9) DYNDH1723, Catalog Number: DYNDH1723; 10) DYNDH1740B, Catalog Number: DYNDH1740B; 11) DYNDH1741B, Catalog Number: DYNDH1741B; 12) DYNDH1771A, Catalog Number: DYNDH1771A; 13) DYNDH1908, Catalog Number: DYNDH1908; 14) DYNDH1940, Catalog Number: DYNDH1940; 15) DYNDH1988, Catalog Number: DYNDH1988; 16) DYNDH2002, Catalog Number: DYNDH2002; 17) DYNDH2017, Catalog Number: DYNDH2017. Blood and urine collection kits

Code / lot info

1) DYNDB3086D, UDI-DI: 10195327313234(each), 40195327313235(case), Lot Number: 24IBA685; 2) DYNDB3100, UDI-DI: 10198459028915(each), 40198459028916(case), Lot Number: 24IBF913; 3) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24FBH708; 4) DYNDH1619A, UDI-DI: 10195327215835(each), 40195327215836(case), Lot Number: 24HBG622; 5) DYNDH1620A, UDI-DI: 10195327216047(each), 40195327216048(case), Lot Number: 24GBY827; 6) DYNDH1620A, UDI-DI: 10195327216047(each), 40195327216048(case), Lot Number: 24HBQ593; 7) DYNDH1620A, UDI-DI: 10195327216047(each), 40195327216048(case), Lot Number: 24HBU881; 8) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24EBI901; 9) DYNDH1694A, UDI-DI: 10195327139957(each), 40195327139958(case), Lot Number: 24HBG834; 10) DYNDH1695, UDI-DI: 10195327113155(each), 40195327113156(case), Lot Number: 24FBQ907; 11) DYNDH1708A, UDI-DI: 10195327498368(each), 40195327498369(case), Lot Number: 24IBQ580; 12) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 24EBI892; 13) DYNDH1714A, UDI-DI: 10195327347550(each), 40195327347551(case), Lot Number: 24IBE380; 14) DYNDH1723, UDI-DI: 10195327134204(each), 40195327134205(case), Lot Number: 24EBI893; 15) DYNDH1723, UDI-DI: 10195327134204(each), 40195327134205(case), Lot Number: 24GBH135; 16) DYNDH1723, UDI-DI: 10195327134204(each), 40195327134205(case), Lot Number: 24HBG826; 17) DYNDH1740B, UDI-DI: 10198459006302(each), 40198459006303(case), Lot Number: 24EBV834; 18) DYNDH1740B, UDI-DI: 10198459006302(each), 40198459006303(case), Lot Number: 24FBH710; 19) DYNDH1740B, UDI-DI: 10198459006302(each), 40198459006303(case), Lot Number: 24FBK197; 20) DYNDH1740B, UDI-DI: 10198459006302(each), 40198459006303(case), Lot Number: 24GBW340; 21) DYNDH1741B, UDI-DI: 10198459006319(each), 40198459006310(case), Lot Number: 24FBJ622; 22) DYNDH1741B, UDI-DI: 10198459006319(each), 40198459006310(case), Lot Number: 24GBW330; 23) DYNDH1771A, UDI-DI: 10198459088971(each), 40198459088972(case), Lot Number: 24IBJ998; 24) DYNDH1908, UDI-DI: 10195327472467(each), 40195327472468(case), Lot Number: 24EBC055; 25) DYNDH1940, UDI-DI: 10195327526535(each), 40195327526536(case), Lot Number: 24FMH509; 26) DYNDH1988, UDI-DI: 10195327628772(each), 40195327628773(case), Lot Number: 24HBF080; 27) DYNDH2002, UDI-DI: 10195327689971(each), 40195327689972(case), Lot Number: 24GBO301; 28) DYNDH2002, UDI-DI: 10195327689971(each), 40195327689972(case), Lot Number: 24HBQ827; 29) DYNDH2017, UDI-DI: 10198459022159(each), 40198459022150(case), Lot Number: 24HBO219