Medline Industries, LP
What this record is not
A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Medline Convenience Kits: 1) PICC CATHETER INSERTION TRAY, Model Number: CVI4300; 2) OSC SHERMAN KNEE ARTHRO BASIC, Model Number: DYKM1977C; 3) KIT ACES INSERT PEG, Model Number: DYKMBNDL130; 4) KIT ACES INSERT PEG, Model Number: DYKMBNDL130A; 5) KIT ACES INSERT PERCUTANEOUS T, Model Number: DYKMBNDL4; 6) NICU LINE CHANGE TRAY, Model Number: LC295
1) CVI4300, UDI-DI: 10653160294313(each), 653160294316(case), Lot Number: 2020120950; 2) DYKM1977C, UDI-DI: 10653160381105(each), 653160381108(case), Lot Number: 2024110190; 3) DYKMBNDL130, UDI-DI: 10193489479027(each), 40193489479028(case), Lot Number: 21ABC197; 4) DYKMBNDL130, UDI-DI: 10193489479027(each), 40193489479028(case), Lot Number: 21ABJ543; 5) DYKMBNDL130, UDI-DI: 10193489479027(each), 40193489479028(case), Lot Number: 21BBB580; 6) DYKMBNDL130A, UDI-DI: 10193489829594(each), 40193489829595(case), Lot Number: 21CLA391; 7) DYKMBNDL130A, UDI-DI: 10193489829594(each), 40193489829595(case), Lot Number: 21CLA391Z; 8) DYKMBNDL130A, UDI-DI: 10193489829594(each), 40193489829595(case), Lot Number: 21CLA621; 9) DYKMBNDL130A, UDI-DI: 10193489829594(each), 40193489829595(case), Lot Number: 21DLA326; 10) DYKMBNDL130A, UDI-DI: 10193489829594(each), 40193489829595(case), Lot Number: 21ELA080; 11) DYKMBNDL130A, UDI-DI: 10193489829594(each), 40193489829595(case), Lot Number: 21FLA214; 12) DYKMBNDL130A, UDI-DI: 10193489829594(each), 40193489829595(case), Lot Number: 21GLA677; 13) DYKMBNDL130A, UDI-DI: 10193489829594(each), 40193489829595(case), Lot Number: 21HLA740; 14) DYKMBNDL130A, UDI-DI: 10193489829594(each), 40193489829595(case), Lot Number: 21ILA163; 15) DYKMBNDL130A, UDI-DI: 10193489829594(each), 40193489829595(case), Lot Number: 21JLA910; 16) DYKMBNDL130A, UDI-DI: 10193489829594(each), 40193489829595(case), Lot Number: 22ALA593; 17) DYKMBNDL130A, UDI-DI: 10193489829594(each), 40193489829595(case), Lot Number: 22BLA721; 18) DYKMBNDL130A, UDI-DI: 10193489829594(each), 40193489829595(case), Lot Number: 22CLA448; 19) DYKMBNDL130A, UDI-DI: 10193489829594(each), 40193489829595(case), Lot Number: 22DLA407; 20) DYKMBNDL130A, UDI-DI: 10193489829594(each), 40193489829595(case), Lot Number: 22ELA008; 21) DYKMBNDL130A, UDI-DI: 10193489829594(each), 40193489829595(case), Lot Number: 22FLA745; 22) DYKMBNDL130A, UDI-DI: 10193489829594(each), 40193489829595(case), Lot Number: 22GLA694; 23) DYKMBNDL130A, UDI-DI: 10193489829594(each), 40193489829595(case), Lot Number: 22GLB021; 24) DYKMBNDL4, UDI-DI: 10193489477627(each), 40193489477628(case), Lot Number: 21ALA056; 25) DYKMBNDL4, UDI-DI: 10193489477627(each), 40193489477628(case), Lot Number: 21ALA631; 26) DYKMBNDL4, UDI-DI: 10193489477627(each), 40193489477628(case), Lot Number: 21DLA343; 27) LC295, UDI-DI: 10653160322702(each), 653160322705(case), Lot Number: 2025110490
Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →