← FDA ENFORCEMENT
Z-2028-2026Class II

Medline Industries, LP

Recalling firm
Medline Industries, LP
Product type
Device
Status
Ongoing
Report date
2026-05-06
Distribution
Worldwide distribution.
Firm state
IL
Product quantity
3000 kits
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Environmental Monitoring
Reason for recall

During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.

Product description

Centurion medical procedure kits, containing Medline Neuro Sponges, labeled as follows: SURGICAL ENT PREP KIT, Medline Kit Number/SKU MNS11715

Code / lot info

UDI/DI EACH 10653160323068, UDI/DI case 40653160323069. Lot Numbers: 23JBY390, 23IBS518, 23FBW067, 23CBO045, 22HBC645, 22GBL079, 22DBP674, 22ABU694, 21IBM298, 21IBM495, 21GBB532, 21EBR576, 21CBL756, 21ABI942, 26ABG606, 25LBS129, 25IBK286, 25GBB222, 25DBH934, 25BBD132, 24LBO080, 24JBW176, 24HBT740, 24EBU727, 24ABW463.

Other recalls SPEQ maps to Environmental Monitoring
2026-07-22Class IFrutas y Hortalizas del Sur S.A.2026-07-22Class IIGELLERT GLOBAL GROUP2026-07-22Class IIChiesi USA, Inc.2026-07-22Class IICareFusion 213, LLC