Heraeus Medical GmbH (Dental Division)
What this record is not
A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.
Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).
Heraeus, PALAMIX uno. Material Number: 66057893.
Material Number: 66057893; UDI-DI: 04260102133587; Batch: all batches with numbers below 223060;