← FDA ENFORCEMENT
Z-1657-2026Class II

Siemens Healthcare Diagnostics, Inc.

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product type
Device
Status
Ongoing
Report date
2026-04-08
Distribution
Domestic: Nationwide Distribution International: Argentina, Australia, Bangladesh, Brazil, Bulgaria , Canada, Chile, China, Colombia, Croatia , Cura¿ao, St Eus, Czech Republic , Denmark , Ecuador, Egypt , Estonia , Finland , France , Georgia , Germany , Greece , Hong Kong, India, Iraq , Italy , Jordan , Kuwait , Latvia , Lithuania , Macedonia , Malaysia, Maldives, Mexico, Netherlands , New Zealand, Norway , Oman , Pakistan , Paraguay, Qatar , Romania , Saudi Arabia , Serbia , Singapore, Slovakia , Slovenia , South Africa, South Korea, Spain , Turkey , Turkmenistan , U.A.E. , Uganda , United Kingdom , Uruguay , Vatikancity;
Firm state
NY
Product quantity
21,291 units (9750 US, 11541 OUS)
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Validation & Qualification
Reason for recall

A potential for depressed results for Atellica CH A1c_E when processing Atellica CH RCRP on the same analyzer. Customers processing A1c_E and RCRP on separate analyzers are not impacted by the recall. The frequency and magnitude of the negative bias and imprecision increase over time and are not consistent with the assay Instructions For Use (IFU). When this issue is present, Quality Control (QC) results demonstrate a negative trend and increased imprecision. The clinical impact of a negative (A1c_E) bias may lead to a delay in diagnosis for diabetes or lead to less-than-optimal treatment for existing diabetes.

Product description

Atellica CH Enzymatic Hemoglobin A1c (A1c_E); Siemens Material Number (SMN): 11097536;

Code / lot info

Siemens Material Number (SMN): 11097536; UDI-DI: 00630414220505; Lot numbers: 150131, 150132, 150232, 150260, 150283, 150344, 150361, 150365, 150366, 150445, 150453, 150460, 150498, 150505, 150562, 150600, 150618, 150621;

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →