Pivotal Health Solutions, Inc.
What this record is not
A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.
The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and melting of the vinyl cushion.
Quantum Intersegmental Table, Model/Catalog Number: Q400; heated massage table
All serial numbers.