← FDA ENFORCEMENT
Z-0186-2026Class II

American Contract Systems Inc.

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
American Contract Systems Inc.
Product type
Device
Status
Ongoing
Report date
2025-10-22
Distribution
US Nationwide distribution in the states of SD, IA, MN, WA, IL.
Firm state
IA
Product quantity
32,433 total
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Environmental Monitoring
Reason for recall

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Product description

Medical convenience kits UDPG88W PICC G-TUBE PACK-230209

Code / lot info

UDPG88W UDI-DI 191072216359 lot 7935611 bag serial number N/A * If bag serial number is listed as "N/A", then whole lot is affected