← FDA ENFORCEMENT
Z-0173-2026Class II

American Contract Systems Inc.

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
American Contract Systems Inc.
Product type
Device
Status
Ongoing
Report date
2025-10-22
Distribution
US Nationwide distribution in the states of SD, IA, MN, WA, IL.
Firm state
IA
Product quantity
32,433 total
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Environmental Monitoring
Reason for recall

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Product description

Medical convenience kits Item number/ Product description AHLD15X LABOR DELIVERY PACK - 213387 AKDV96B VAGINAL DELIVERY AKDV96B VAGINAL DELIVERY EVBH28D BIRTHING ROOM PACK UPVD01K VAGINAL DELIVERY PACK

Code / lot info

AHLD15X UDI-DI 191072224767 Lot 8188711 Bag serial no. 17555039 17555044 17555049 17609166 17609181; AKDV96B UDI-DI 191072224019 Lot 7311311 Bag serial no. 47140875 47141919; EVBH28D UDI-DI 191072239051 Lot 8424611 Bag serial no. 17754395; UPVD01K UDI-DI 191072238993 Lot 8439111 Bag serial no. 17682598 Log 8499211 Bag serial no. 17742543

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →