← FDA ENFORCEMENT
Z-0167-2027Class II

Howmedica Osteonics Corp.

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Howmedica Osteonics Corp.
Product type
Device
Status
Ongoing
Report date
2026-10-21
Distribution
These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.
Firm state
NJ
Product quantity
577 trays
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Environmental Monitoring
Reason for recall

Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.

Product description

TS Misc. Instruments Tray, Triathlon. Catalog Number: 6543-6-519.

Code / lot info

Catalog Number: 6543-6-519. GTIN: 07613327555127. All lot numbers used prior to the correction date of May 8, 2026 are impacted.

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →