← FDA ENFORCEMENT
Z-0072-2027Class II

Steris Corporation

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Steris Corporation
Product type
Device
Status
Ongoing
Report date
2026-10-14
Distribution
Worldwide - US Nationwide distribution.
Firm state
OH
Product quantity
1185
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Environmental Monitoring
Reason for recall

Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

Product description

VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3060

Code / lot info

UDI: 10724995023833 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027 270302: expiration date - 03/02/2027