← FDA ENFORCEMENT
Z-0045-2027Class II

ASP Global, LLC.

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
ASP Global, LLC.
Product type
Device
Status
Ongoing
Report date
2026-10-07
Distribution
US Nationwide distribution in the states of MO, NV, IN, MN, GA, WI, PA, TX, CT, LA, NJ, FL, AL, NY, SC, NE, NC, OH, AZ, CO, MA, IL, KY, CA, VA, ME, KS, MD, TN, MS, AR, WA, NM, MI, WY.
Firm state
GA
Product quantity
53040
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Validation & Qualification
Reason for recall

Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.

Product description

Brand Name: ASP Global Product Name:DISPOSABLE TOURNIQUET CUFF, 34" X 4" 1-Line, Blue Model/Catalog Number: ASP-564012 Software Version: Not Applicable Product Description: Sterile Disposable Tourniquet Cuff is used to be wrapped around a patient's limb and inflated to reduce or totally occlude circulation during surgery. It is non-invasive device. Tourniquet Cuffs are intended for use in surgical procedures that requires the interruption of blood flow and clear visualization of the operative field. Component: No

Code / lot info

Lot Code: Model Number: ASP-564012 UDI-DI: (01) 00840216106772 All Lots produced from 01SEP2016 to 01MAR2023: P1015STY, P1215STY, O0815STY, O1020STY, N0330STY, N0530STY, N0730STY, M0130STY, M0430STY, M0730STY, M0930STY, M1015STY, M1101STY, M1115STY, L0310STY, L0325STY