FDA PRODUCT CODE · CLASS II
PTL — Wire, Guide, Catheter, Exempt
This product code is the class II exempt counterpart of DQX, and was exempted, subject to the limitations of exemption, under the procedures established by the 21st Century Cures Act. This device type does not require premarket notification (510(k)) to market if it does not exceed the general limitations of exemption found in 21 CFR 870.9 and the specific limitations of exemption as stated in the Physical State field of this product code description.
What these counts are not
Clearance and recall counts are SPEQ’s aggregation of FDA’s own published records for the code. A recall count is a measure of history, not of a device’s safety, neither number is an FDA rating, and SPEQ is not affiliated with the FDA.
What this classification requires
Every statement below is FDA’s own classification record for this product code, quoted from the device classification field reference. The SPEQ reading beneath each one is SPEQ’s interpretation of what it means for the work, and is not FDA text.
510(K) Exempt
SPEQ READINGNo premarket notification is required for a device in this code, within the limitations of exemption. Exempt from premarket review is NOT exempt from regulation: registration, listing, labelling, adverse-event reporting and — unless the code is also GMP exempt — 21 CFR Part 820 all still apply.
Class II (moderate to high risk): general controls and special controls
SPEQ READINGSpecial controls are the distinguishing feature: performance standards, guidance documents, patient registries or postmarket surveillance named for the device type. Find the special controls before designing the test plan — they usually name the tests.
Device is eligible for The Voluntary Malfunction Summary Reporting Program
SPEQ READINGMalfunctions for this code may be reported in periodic summary form rather than as individual MDRs under 21 CFR Part 803. Eligibility is a property of the product code; using it is a decision the manufacturer makes and must be able to justify.
Not exempt due to Good Manufacturing Practice (GMP)/Quality System
SPEQ READINGThe full Quality Management System Regulation (21 CFR 820, incorporating ISO 13485:2016) applies, and is what an FDA inspection of this device will be conducted against.
Device is not implantable
SPEQ READINGNo implant-specific expectations attach from the classification itself.
Device is not used for life sustaining purposes
SPEQ READINGNo life-sustaining designation attaches to this code.
Device is not a candidate for a third party review program
SPEQ READINGFDA reviews any premarket notification for this code directly.
Other product codes under 21 CFR 870.1330
One regulation usually governs several product codes. These share this code’s regulation, and therefore its classification basis.
21 CFR 870.1330 — Catheter guide wire
The classification regulation for this product code, as published by the Office of the Federal Register. This is the regulation text itself — no SPEQ interpretation appears in it.
(a) Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
(b) Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9. [45 FR 7907, Feb. 5, 1980, as amended at 84 FR 71811, Dec. 30, 2019]
Source: Electronic Code of Federal Regulations, Title 21. 21 CFR is a work of the United States Government and is in the public domain. SPEQ read this section on October 21, 2026. CFR is amended continuously, so verify against the source before acting on it: read 21 CFR 870.1330 on eCFR →
Source: FDA device classification, 510(k)/PMA, and recall datasets. Recall entries count distinct FDA recall numbers, so one recall event affecting several products appears as several entries. Browse every product code SPEQ tracks →