FDA PRODUCT CODE · CLASS II

DQX — Wire, Guide, Catheter

What these counts are not

Clearance and recall counts are SPEQ’s aggregation of FDA’s own published records for the code. A recall count is a measure of history, not of a device’s safety, neither number is an FDA rating, and SPEQ is not affiliated with the FDA.

DEVICE CLASS
Class II
REGULATION
21 CFR 870.1330
MEDICAL SPECIALTY
Cardiovascular
REVIEW PANEL
CV
— Implant— Life-sustaining— GMP-exempt✓ Third-party review eligible
26
510(K)/PMA CLEARANCES
510k: 26
84
RECALL ENTRIES

What this classification requires

Every statement below is FDA’s own classification record for this product code, quoted from the device classification field reference. The SPEQ reading beneath each one is SPEQ’s interpretation of what it means for the work, and is not FDA text.

PREMARKET PATHWAY

510(K)

SPEQ READINGA premarket notification under 21 CFR Part 807 Subpart E is the route for this code. The submission argues substantial equivalence to a legally marketed predicate, so the engineering question is which predicate the design can actually be compared against, and the evidence question is what testing supports that comparison.

DEVICE CLASS

Class II (moderate to high risk): general controls and special controls

SPEQ READINGSpecial controls are the distinguishing feature: performance standards, guidance documents, patient registries or postmarket surveillance named for the device type. Find the special controls before designing the test plan — they usually name the tests.

MALFUNCTION REPORTING

Device is eligible for The Voluntary Malfunction Summary Reporting Program

SPEQ READINGMalfunctions for this code may be reported in periodic summary form rather than as individual MDRs under 21 CFR Part 803. Eligibility is a property of the product code; using it is a decision the manufacturer makes and must be able to justify.

GMP / QUALITY SYSTEM

Not exempt due to Good Manufacturing Practice (GMP)/Quality System

SPEQ READINGThe full Quality Management System Regulation (21 CFR 820, incorporating ISO 13485:2016) applies, and is what an FDA inspection of this device will be conducted against.

IMPLANT

Device is not implantable

SPEQ READINGNo implant-specific expectations attach from the classification itself.

LIFE-SUSTAINING

Device is not used for life sustaining purposes

SPEQ READINGNo life-sustaining designation attaches to this code.

THIRD-PARTY REVIEW

Device is a candidate for a third party review program

SPEQ READINGAn Accredited Person may conduct the 510(k) review instead of FDA. FDA still makes the clearance decision — third-party review changes who reviews, never who decides.

Other product codes under 21 CFR 870.1330

One regulation usually governs several product codes. These share this code’s regulation, and therefore its classification basis.

21 CFR 870.1330 — Catheter guide wire

The classification regulation for this product code, as published by the Office of the Federal Register. This is the regulation text itself — no SPEQ interpretation appears in it.

(a) Identification. A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

(b) Classification. Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9. [45 FR 7907, Feb. 5, 1980, as amended at 84 FR 71811, Dec. 30, 2019]

Source: Electronic Code of Federal Regulations, Title 21. 21 CFR is a work of the United States Government and is in the public domain. SPEQ read this section on October 21, 2026. CFR is amended continuously, so verify against the source before acting on it: read 21 CFR 870.1330 on eCFR →

Recent clearances

K253146Rapid-Medical , Ltd. · 510k2026-09-18
K261281Filmecc Co., Ltd. · 510k2026-09-03
K253667Sp Medical A/S · 510k2026-07-14
K261640Merit Medical Ireland , Ltd. · 510k2026-07-13
K254137Merit Medical Ireland, Ltd. · 510k2026-05-22
K260544FMD Co., Ltd. · 510k2026-03-20
K253746Lake Region Medical · 510k2026-03-19
K253262Nano4imaging GmbH · 510k2026-03-06

Recent recalls

Cook IncorporatedOpen, Classified · 2024-11-21

Devices from the affected device lots have labels that state the incorrect expiration dates.

Cook IncorporatedOpen, Classified · 2024-11-08

Affected device lots have labels that state the incorrect expiration dates.

Cook IncorporatedOpen, Classified · 2024-11-08

Affected device lots have labels that state the incorrect expiration dates.

Cook Inc.Terminated · 2020-11-03

Affected product was dropped during manufacturing, which may result in a bent wire guide and/or the presence of loose foreign matter on the wire guide.

Stryker NeurovascularTerminated · 2019-10-29

The product labels (pouch and carton) for products contain an incorrect value in millimeters (0.036mm) for the Guidewire Outer Diameter dimension. The inches dimension (0.014in) on the label is correct. The correct values are 0.014in (0.36mm). 0.036mm represents a typographical error.

Stryker NeurovascularTerminated · 2019-10-29

The product labels (pouch and carton) for products contain an incorrect value in millimeters (0.036mm) for the Guidewire Outer Diameter dimension. The inches dimension (0.014in) on the label is correct. The correct values are 0.014in (0.36mm). 0.036mm represents a typographical error.

Stryker NeurovascularTerminated · 2019-10-29

The product labels (pouch and carton) for products contain an incorrect value in millimeters (0.036mm) for the Guidewire Outer Diameter dimension. The inches dimension (0.014in) on the label is correct. The correct values are 0.014in (0.36mm). 0.036mm represents a typographical error.

Stryker NeurovascularTerminated · 2019-10-29

The product labels (pouch and carton) for products contain an incorrect value in millimeters (0.036mm) for the Guidewire Outer Diameter dimension. The inches dimension (0.014in) on the label is correct. The correct values are 0.014in (0.36mm). 0.036mm represents a typographical error.

Source: FDA device classification, 510(k)/PMA, and recall datasets. Recall entries count distinct FDA recall numbers, so one recall event affecting several products appears as several entries. Browse every product code SPEQ tracks →