FDA PRODUCT CODE · CLASS II
FDE — Laparoscopy Kit
This product code has been established in accordance with the May 20, 1997, Guidance entitled, Convenience Kits Interim Regulatory Guidance, found at www.fda.gov/cdrh/ode/convkit.html. This type of convenience kit, as listed in the guidance above, is under enforcement discretion, and does not require a premarket notification (510(k)) to market if it meets all criteria in the guidance.
What these counts are not
Clearance and recall counts are SPEQ’s aggregation of FDA’s own published records for the code. A recall count is a measure of history, not of a device’s safety, neither number is an FDA rating, and SPEQ is not affiliated with the FDA.
What this classification requires
Every statement below is FDA’s own classification record for this product code, quoted from the device classification field reference. The SPEQ reading beneath each one is SPEQ’s interpretation of what it means for the work, and is not FDA text.
FDA publishes no definition for this value in the device classification field reference, so SPEQ does not interpret it.
Class II (moderate to high risk): general controls and special controls
SPEQ READINGSpecial controls are the distinguishing feature: performance standards, guidance documents, patient registries or postmarket surveillance named for the device type. Find the special controls before designing the test plan — they usually name the tests.
Device is eligible for The Voluntary Malfunction Summary Reporting Program
SPEQ READINGMalfunctions for this code may be reported in periodic summary form rather than as individual MDRs under 21 CFR Part 803. Eligibility is a property of the product code; using it is a decision the manufacturer makes and must be able to justify.
Not exempt due to Good Manufacturing Practice (GMP)/Quality System
SPEQ READINGThe full Quality Management System Regulation (21 CFR 820, incorporating ISO 13485:2016) applies, and is what an FDA inspection of this device will be conducted against.
Device is not implantable
SPEQ READINGNo implant-specific expectations attach from the classification itself.
Device is not used for life sustaining purposes
SPEQ READINGNo life-sustaining designation attaches to this code.
Device is not a candidate for a third party review program
SPEQ READINGFDA reviews any premarket notification for this code directly.
Other product codes under 21 CFR 876.1500
One regulation usually governs several product codes. These share this code’s regulation, and therefore its classification basis.
21 CFR 876.1500 — Endoscope and accessories
The classification regulation for this product code, as published by the Office of the Federal Register. This is the regulation text itself — no SPEQ interpretation appears in it.
(a) Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
(b) Classification —(1) Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9. [48 FR 53023, Nov. 23, 1983, as amended at 61 FR 1122, Jan. 16, 1996; 66 FR 38801, July 25, 2001; 83 FR 25914, June 5, 2018; 84 FR 71813, Dec. 30, 2019]
Source: Electronic Code of Federal Regulations, Title 21. 21 CFR is a work of the United States Government and is in the public domain. SPEQ read this section on October 22, 2026. CFR is amended continuously, so verify against the source before acting on it: read 21 CFR 876.1500 on eCFR →
Recent recalls
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.
Laparoscopy packs labeled as "latex free" were found to contain a latex urethral catheter
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to
Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness can
Custom convenience kits are being recall because the contain non-sterile Codman Surgical Patties & Surgical Strips, which have the potential for endotoxin contamination.
Surgical Patties & Surgical Strips in specific lots of Custom Convenience Kits were found to contain higher-than-expected levels of endotoxin within the raw material used to produce sterile Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in those
Source: FDA device classification, 510(k)/PMA, and recall datasets. Recall entries count distinct FDA recall numbers, so one recall event affecting several products appears as several entries. Browse every product code SPEQ tracks →