FDA PRODUCT CODE · CLASS II
FBN — Choledochoscope And Accessories, Flexible/Rigid
to examine and perform procedures within the bile ducts. For reusable, flexible endoscopes in this product code, validated reprocessing instructions and reprocessing validation data must be included in a 510(k) submission (82 FR 26807, available at https://www.gpo.gov/fdsys/pkg/FR-2017-06-09/pdf/2017-12007.pdf).
What these counts are not
Clearance and recall counts are SPEQ’s aggregation of FDA’s own published records for the code. A recall count is a measure of history, not of a device’s safety, neither number is an FDA rating, and SPEQ is not affiliated with the FDA.
What this classification requires
Every statement below is FDA’s own classification record for this product code, quoted from the device classification field reference. The SPEQ reading beneath each one is SPEQ’s interpretation of what it means for the work, and is not FDA text.
510(K)
SPEQ READINGA premarket notification under 21 CFR Part 807 Subpart E is the route for this code. The submission argues substantial equivalence to a legally marketed predicate, so the engineering question is which predicate the design can actually be compared against, and the evidence question is what testing supports that comparison.
Class II (moderate to high risk): general controls and special controls
SPEQ READINGSpecial controls are the distinguishing feature: performance standards, guidance documents, patient registries or postmarket surveillance named for the device type. Find the special controls before designing the test plan — they usually name the tests.
Device is not eligible for The Voluntary Malfunction Summary Reporting Program
SPEQ READINGIndividual malfunction reporting under 21 CFR Part 803 applies. The complaint-handling system has to produce a decision per event, not per period.
Not exempt due to Good Manufacturing Practice (GMP)/Quality System
SPEQ READINGThe full Quality Management System Regulation (21 CFR 820, incorporating ISO 13485:2016) applies, and is what an FDA inspection of this device will be conducted against.
Device is not implantable
SPEQ READINGNo implant-specific expectations attach from the classification itself.
Device is not used for life sustaining purposes
SPEQ READINGNo life-sustaining designation attaches to this code.
Device is not a candidate for a third party review program
SPEQ READINGFDA reviews any premarket notification for this code directly.
Other product codes under 21 CFR 876.1500
One regulation usually governs several product codes. These share this code’s regulation, and therefore its classification basis.
21 CFR 876.1500 — Endoscope and accessories
The classification regulation for this product code, as published by the Office of the Federal Register. This is the regulation text itself — no SPEQ interpretation appears in it.
(a) Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
(b) Classification —(1) Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9. [48 FR 53023, Nov. 23, 1983, as amended at 61 FR 1122, Jan. 16, 1996; 66 FR 38801, July 25, 2001; 83 FR 25914, June 5, 2018; 84 FR 71813, Dec. 30, 2019]
Source: Electronic Code of Federal Regulations, Title 21. 21 CFR is a work of the United States Government and is in the public domain. SPEQ read this section on October 22, 2026. CFR is amended continuously, so verify against the source before acting on it: read 21 CFR 876.1500 on eCFR →
Recent clearances
Recent recalls
Potential for inadvertent retention of biomaterial when the reprocessing procedure is not performed as instructed, so as not completely removing and disassembling the device.
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
The CHF-CB30S Device did not have a 510(k) clearance. Additionally, the device has been associated with breakage of the insertion section and protrusion of metal parts, as well as breakage and displacement of the rubber on the bending section during surgical procedures.
Source: FDA device classification, 510(k)/PMA, and recall datasets. Recall entries count distinct FDA recall numbers, so one recall event affecting several products appears as several entries. Browse every product code SPEQ tracks →