Philips North America
What this record is not
A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.
Ingenia 1.5T, 1.5T CX, and 1.5T S; Model number (REF): 1. 781315 (1.5T), 2. 781341 (1.5T), 3. 781396 (1.5T), 4. 782115 (1.5T), 5. 782140 (1.5T), 6. 781262 (1.5T CX), 7. 781347 (1.5T S); Software Versions: R11, R12 and R13;
1. Model number (REF): 1. 781315 (1.5T); UDI-DI: 00884838099715; Serial Numbers: 61005, 61007, 61014, 61033, 61036, 61047, 61056, 61057, 61076, 61087, 61088, 61105, 61116; 2. Model Number (REF): 781341 (1.5T); UDI-DI: 00884838055322; Serial Numbers: 70042, 70051, 70056, 70062, 70203, 70227, 70229, 70249, 70257, 70262, 70285, 70300, 70311, 70383, 70424, 70432, 70438, 70445, 70453, 70460, 70472, 70588, 70598, 70619, 70638, 70642, 70645, 70647, 70708, 70717, 70724, 70728, 70743, 70750, 70751, 70787, 70801, 70821, 70825, 70836, 70846, 70875, 70876, 70886, 70889, 70896, 70904, 70920, 70957, 70961, 70967, 70970, 70979, 70988, 84049, 84076, 84079, 84080, 84141, 84162, 84170, 84173, 84186, 84205, 84212, 84219, 84233, 84250, 84254, 84266, 84267, 84281, 84289, 84329, 84332, 84335, 84343, 84359, 84360, 84368, 84379, 84388, 84392, 84403, 84415, 84431, 84433, 84436, 84450, 84451, 84467, 84476, 84491, 84500, 84510, 84539, 84549, 84555, 84556, 84568, 84596, 84634, 84635, 84641, 84642, 84652, 84660, 84708, 84713, 84714, 84721, 84741, 84745, 84768, 84804, 84839, 84840, 84853, 84858, 84867, 84869, 84893, 84904, 84908, 84978, 84987, 87068, 87076, 87092, 87196, 87197, 87259; 3. Model Number (REF): 781396 (1.5T); UDI-DI: 00884838009820; Serial Numbers: 41000, 41001, 41021, 41034, 41046, 41056, 41061, 41091, 41116, 41145, 41194, 41231, 41249, 41253, 41254, 41255, 41352, 41354, 41356, 41370, 41400, 41408, 41422, 41428, 41456, 41475, 41492, 41513, 41515, 41548; 4. Model Number (REF): 782115 (1.5T); UDI-DI: 00884838099043; Serial Numbers: 61186, 61191, 61196, 61219, 61228, 61236, 61238, 61272, 61276, 61317, 61382, 61383, 61384, 61386, 61387, 61388, 61390, 61396, 61399, 61401, 61402, 61408, 61411, 61418, 61430; 5. Model Number (REF): 782140 (1.5T); UDI-DI: 00884838108646; Serial Numbers: 45608, 45609; 6. Model Number (REF): 781262 (1.5T CX); UDI-DI: 00884838068445; Serial Numbers: 77047, 77058, 77059, 77120, 77143, 77151, 77158, 77161, 77179, 77180, 77182; 7. Model Number (REF): 781347 (1.5T S); UDI-DI: 00884838068421; Serial Numbers: 81027, 81033, 81145;
Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →