← FDA ENFORCEMENT
Z-3249-2026Class II

Philips Ultrasound, LLC

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Philips Ultrasound, LLC
Product type
Device
Status
Ongoing
Report date
2026-09-30
Distribution
Worldwide - US Nationwide distribution in the states of AL, CA, CT, FL, GA, KS, MD, MN, NJ, NY, OH, PA, SC, TN, TX, VA, WA, WI and the countries of Austria, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lithuania, Netherlands, Norway, Philippines, Poland, Romania, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan , United Kingdom, Vietnam.
Firm state
WA
Product quantity
219 units
Initiated by
Voluntary: Firm initiated
Reason for recall

Due to software issue with ultrasound system, doppler sample is displayed shallower than its actual measurement location, potentially causing inaccurate blood flow velocity measurements and misdiagnosis.

Product description

Philips EPIQ Elite Diagnostic Ultrasound System Model # 795234

Code / lot info

All Philips EPIQ Elite Ultrasound Systems running software version 14.0, UDI(s): (01)00884838102514 (01)00884838107540 (01)00884838114159 (01)00884838124912 (01)00884838128125