← FDA ENFORCEMENT
Z-3231-2026Class II

Biomerieux Inc

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Biomerieux Inc
Product type
Device
Status
Ongoing
Report date
2026-09-30
Distribution
International distribution to the countries of Austria, Germany, Luxembourg.
Firm state
MO
Product quantity
97680 units
Initiated by
Voluntary: Firm initiated
Reason for recall

The issue is related to card pouch damage associated with a portion of the lot which was produced on a specific card pouching device, in a specific timeframe, causing a potential risk of false resistant results. False resistant results may lead to the use of less effective antibiotic treatments or the association of multiple antibiotics having negative side effects, which could potentially impact patient s health.

Product description

VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811

Code / lot info

UDI-DI 03573026637712; Lot Number 1333393504