← FDA ENFORCEMENT
Z-3193-2026Class II

Mirion Technologies (Capintec), Inc.

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Mirion Technologies (Capintec), Inc.
Product type
Device
Status
Ongoing
Report date
2026-09-23
Distribution
Worldwide distribution - US Nationwide and the countries of ALGERIA, AUSTRIA, BANGLADESH, BRASIL, Brazil, CANADA, CHILE, Costa Rica, Ecuador, France, Hanoi Vietnam, Hong Kong, INDIA, ISRAEL, ITALY, Malaysia, Philippines, POLAND, Saudi Arabia, SOUTH KOREA, Spain, SWEDEN, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIATES, UNITED KINGDOM, VIETNAM.
Firm state
NJ
Product quantity
783
Initiated by
Voluntary: Firm initiated
Reason for recall

Potential for failure of the tension bolt which would allow the spring arm to unexpectedly fall to its lowest position, which is approximately 25 inches (63.5 cm) from the ground.

Product description

Captus 4000e Thyroid Uptake System, Models: 1. 5430-30151, 2. 5430-30152, 3. 5430-30154;

Code / lot info

Affected Units: Captus 4000: 1. Model 5430-30151; UDI-DI: 00859942006096; 2. Model 5430-30152; UDI-DI: 00859942006102; 3. Model 5430-30154; UDI-DI: 00859942006423; Devices shipped from April 2019 through February 2024, Serial Numbers 940548 through 941398, that have NOT already replaced the original arm with the redesigned arm.

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →