← FDA ENFORCEMENT
Z-3191-2026Class II

Datex-Ohmeda, Inc.

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Datex-Ohmeda, Inc.
Product type
Device
Status
Ongoing
Report date
2026-09-23
Distribution
Worldwide - US Nationwide distribution.
Firm state
WI
Product quantity
7 units
Initiated by
Voluntary: Firm initiated
Reason for recall

Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.

Product description

Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000

Code / lot info

Serial Numbers: ANAS01357 ANAP00609 ANAQ00988 ANAS00586 ANAR00543 ANAP01088 ANAM00443