← FDA ENFORCEMENT
Z-3178-2026Class II

restor3d Inc.

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
restor3d Inc.
Product type
Device
Status
Ongoing
Report date
2026-09-23
Distribution
US Nationwide distribution in the states of FL, CO.
Firm state
NC
Product quantity
2
Initiated by
Voluntary: Firm initiated
Reason for recall

Ankle implantation systems were packaged with incorrect patient-specific cutting guides, which may result in: delayed surgery, the need to use standard instrumentation, reduced fixation or poor alignment of implants, implant loosening/migration, or need for revision surgery.

Product description

Kinos Axiom Total Ankle System with Axiom PSR, Flat-cut (Cut Guide), REF: 5831-9006

Code / lot info

System Serial Numbers: 18169, 18059. REF: 5831-9006, Lot: D-788, UDI-DI: 00840097509945.

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →