Z-3090-2026Class I

Given Imaging Ltd.

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Given Imaging Ltd.
Product type
Device
Status
Ongoing
Report date
2026-09-16
Distribution
Worldwide distribution. US nationwide including Puerto Rico; Argentina, Australia, Austria, Bahrain, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Czech Republic, Dominican Republic, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Kazakhstan, Kenya, Luxembourg, Netherlands, New Zealand, Norway, Panama, Poland, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, United Arab Emirates, and United Kingdom.
Product quantity
127,416
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Risk Management & CAPA
Reason for recall

Capsule may fail to attach to the esophagus or detach from the delivery device. Risks include patient aspiration/inhalation, perforation of esophagus, obstruction of the airway, hemorrhage/blood loss/bleeding, laceration of the esophagus, a delay in diagnosis, and foreign bodies remaining in the patient.

Product description

Bravo CF capsule delivery device, 5-pack, Product Number FGS-0635, Bravo CF capsule delivery device, 1-pack, Product Number FGS-0636. Used to attach a pH monitoring capsule to the esophagus.

Code / lot info

Product Number FGS-0635 UDI-DI (GTIN) 07290101369707 Product Number FGS-0636 UDI-DI (GTIN) 07290101369714 Expiration dates prior to and including October 20, 2027 (lot numbers from 64553F up to and including 68912F)