United Orthopedic Corp.
What this record is not
A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.
Due to an incomplete process of the component screw by a subcontractor which prevents the properly assembling of stem trail.
Brand Name: U2 Total Knee System Product Name: Femoral valgus adaptor, left and right Model/Catalog Number: 9323-5333 and 9323-5334 Software Version: N/A Product Description: Manual Orthopedic Surgical Instrument Component: No
Product Number; UDI-DI Code; Lot Code: Product number 9323-5333; UDI-DI 04719886908084; Lot number 25FB50 Product number 9323-5334; UDI-DI 04719886908091; Lot number 25FB51