← FDA ENFORCEMENT
Z-2877-2026Class II

Biofire Defense

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Biofire Defense
Product type
Device
Status
Ongoing
Report date
2026-08-12
Distribution
U.S. Nationwide distribution in the states of AL, CA, DC, FL, GA, HI, KY, LA, MA, MD, MN, NC, NH, NV, NY, OH, RI, TN, TX, UT, VA and WA.
Firm state
UT
Product quantity
122 kits (3,660 pouches)
Initiated by
Voluntary: Firm initiated
Reason for recall

Firm has identified mismatches between the outbreak strain and the reference strain used in the development of its qualitative, multiplexed, nucleic acid-based in vitro diagnostic panel test.

Product description

FilmArray NGDS Warrior Panel REF NGDS-ASY-0007 UDI-DI code: (01)00851458005136 The FilmArray NGDS Warrior Panel is a qualitative, multiplexed, nucleic acid-based in vitro diagnostic test intended for use with the FilmArray 2.0 system. The FilmArray NGDS Warrior Panel detects and identifies Bacillus anthracis, Yersinia pestis, Francisella tularensis, Coxiella burnetii, Ebola virus, and Marburg virus nucleic acids directly from human whole blood (EDTA).

Code / lot info

Part Number: NGDS-ASY-0007 UDI-DI code: (01)00851458005136 Lot Numbers: 311825D 312025D 313526D 322925D

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →