← FDA ENFORCEMENT
Z-2825-2026Class II

Instrumentation Laboratory

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Instrumentation Laboratory
Product type
Device
Status
Ongoing
Report date
2026-08-05
Distribution
US Nationwide distribution.
Firm state
MA
Product quantity
21
Initiated by
Voluntary: Firm initiated
Reason for recall

Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.

Product description

Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Portable blood analyzer system

Code / lot info

GEM Premier 5000: Part number: 00024019255 UDI: 08426950870227 Units distributed in Spain: GEM Premier 5000 Serial Nos.: 25028719, 18021546, 18021549, 26012036, 18051850, 23054764, 21068342, 25080219, 26010781, 25122012, 25041639, 25120558, 25079944, 25079943, 25028719, 25100485, 25079947, 25100483, 25100484, 21057870 and 26010710. Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634