← FDA ENFORCEMENT
Z-2775-2026Class II

KENT IMAGING, INC.

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
KENT IMAGING, INC.
Product type
Device
Status
Ongoing
Report date
2026-07-29
Distribution
US Nationwide distribution in the states of PA, IA, FL, TX, GA, NV, CA, UT, IN, DC, KS, IL, TN, AR, AZ.
Firm state
N/A
Product quantity
77
Initiated by
Voluntary: Firm initiated
Reason for recall

The software controlling the wound imaging device's battery charging circuit was not designed to recognize and recover a battery that has fully discharged into a low-power "shipping mode." When this occurs, an affected device will not resume charging when reconnected to power and becomes temporarily unusable/ unavailable for wound imaging until corrected.

Product description

SnapshotGLO Wound Imaging Device, Model KB100, KB100AA.

Code / lot info

UDI-DI: B376KB1000, Software Revisions v4.2.0 and v4.1.3