← FDA ENFORCEMENT
Z-2759-2026Class II

ICU Medical Inc.

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
ICU Medical Inc.
Product type
Device
Status
Ongoing
Report date
2026-07-29
Distribution
Worldwide - US Nationwide distribution.
Firm state
MN
Product quantity
39,750 units
Initiated by
Voluntary: Firm initiated
Reason for recall

ICU Medical became aware of an increased likelihood of leakage in specific Lots of L-70 HOTLINE Fluid Warming Sets due to potentially inadequate solvent application at the tubing to return connector bond.

Product description

Hotline Fluid Warming Set, single use device, is designed to warm blood and intravenous fluids and deliver them to the patient's intravenous access site at normothermic temperatures under gravity flow conditions, List Number L-70

Code / lot info

UDI-DI: 30695085407007; Lot Numbers: 4293568, 4293569, 4297510, 4297509, 4326087, 4337724, 4337725, 4346349