Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
Photera 600, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
Serial Numbers: 6-09033NB4CL, 6-09037NB4CL, 6-09031NB4CL, 6-09538NB4CL, 6-09038NB4CL, 6-09039NB4CL, 6-09040NB4CL, 6-09041NB4CL, 6-09035NB4CL, 6-09034NB4CL, 6-09030NB4CL, 6-09026NB4CL, 6-09023NB4CL, 6-09028NB4CL, 6-09025NB4CL, 6-09043NB4CL, 6-09029NB4CL, 6-09044NB4CL, 6-09022NB4CL, 6-09027NB4CL, 6-09036NB4CL, 6-09045NB4CL, 6-09024NB4CL, 6-07827NB4CL, 6-09539NB4CL, 6-09042NB4CL.
Data: U.S. FDA openFDA enforcement API. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification.