← FDA ENFORCEMENT
Z-2716-2026Class II

Think Surgical, Inc.

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Think Surgical, Inc.
Product type
Device
Status
Ongoing
Report date
2026-07-15
Distribution
U.S. Nationwide distribution in the states of AZ, CA, DE, FL, IA, KS, LA, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OR, PA, TX, UT, VA, and WA.
Firm state
CA
Product quantity
360 devices
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Risk Management & CAPA
Reason for recall

Due to manufacturing/design issue where the knob on the tracking array may loosen causing potential misalignment of the array.

Product description

TMINI Femur Tracking Array REF 109003 UDI-DI code: M9461090030 TMINI Tibia Tracking Array REF 109004 UDI-DI code: M9461090040 TMINI Miniature Robotic System is indicated as a stereotaxic instrumentation system for total knee replacement (TKA) surgery. It is to assist the surgeon by providing software-defined spatial boundaries for orientation and reference information to identifiable anatomical structures for the accurate placement of knee implant components

Code / lot info

Model Number: 109003 UDI-DI code: M9461090030 Model Number: 109004 UDI-DI code: M9461090040