Medline Industries, LP
What this record is not
A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Medline Convenience kits containing recalled Swan-Ganz Catheters RIGHT HEART CATH PACK CTX DYNJ68530A
UDI-DI 10195327033620 Lots 21KBN891 21KBO410 22ABA676 22ABY937 22BBN909 22BBT447 23ABO303