Zimmer, Inc.
What this record is not
A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.
Five complaints have been received identifying issues where the plastic guides were catching during reaming, leading to breakage.
Off-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide Model/Catalog Number: 110040241
Lot Code: Model No. 110040241 ; UDI-DI (01)00889024681941(17)350602(10)099463 ; Lot Number 099463 Model No. 110040241 ; UDI-DI (01)00889024681941(17)350531(10)099464 ; Lot Number 099464 Model No. 110040241 ; UDI-DI (01)00889024681941(17)350812(10)099465 ; Lot Number 099465 Model No. 110040241 ; UDI-DI (01)00889024681941(17)350707(10)099467 ; Lot Number 099467
Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →