PFM MEDICAL INC.
What this record is not
A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.
A snare catheter including luer lock, indicated for removal or manipulation of a foreign body in the vascular system, failed to meet the biocompatibility requirements for in-vitro-cytotoxicity, which if in contact with a patient's vasculature could result in localized inflammation and tissue irritation.
Multi-Snare Set: 5 mm x 125 cm, REF: 147305; 10 mm x 125 cm, REF: 147310
REF/UDI-DI/Lot(Expiration): 147305/04042301061829/1049868(10/14/2028); 147310/04042301033567/1049960(10/20/28)