← FDA ENFORCEMENT
Z-2479-2026Class II

Medline Industries, LP

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Medline Industries, LP
Product type
Device
Status
Ongoing
Report date
2026-06-24
Distribution
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
Firm state
IL
Product quantity
2237471 units
Initiated by
Voluntary: Firm initiated
SPEQ quality domain
Validation & Qualification
Reason for recall

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

Product description

MEDLINE Urethral Catheter, Red Rubber Latex, Medline Product Number/SKU (REF), French sizes: 1. DYND13508 - 08 Fr; 2. DYND13510 - 10 Fr; 3. DYND13512 - 12 Fr; 4. DYND13514 - 14 Fr; 5. DYND13515 - 15 Fr; 6. DYND13516 - 16 Fr; 7. DYND13518 - 18 Fr; 8. DYND13520 - 20 Fr; 9. DYND13522 - 22 Fr; 10. DYND13614 - 14 Fr; 11. DYND13616 - 16 Fr.

Code / lot info

1. DYND13508, UDI/DI 10080196031139 (each), 20080196031136 (case); 2. DYND13510, UDI/DI 10080196726462 (each), 40080196726463 (case); 3. DYND13512, UDI/DI 10080196031177 (each), 40080196031178 (case); 4. DYND13514, UDI/DI 10080196726219 (each), 40080196726210 (case); 5. DYND13515, UDI/DI 10080196031214 (each), 40080196031215 (case); 6. DYND13516, UDI/DI 10080196031221 (each), 40080196031222 (case); 7. DYND13518, UDI/DI 10080196031245 (each), 40080196031246 (case); 8. DYND13520, UDI/DI 10080196031252 (each), 40080196031253 (case); 9. DYND13522, UDI/DI 10080196061259 (each), 40080196061250 (case); 10. DYND13614, UDI/DI 10080196031276 (each), 30080196031270 (case); 11. DYND13616, UDI/DI 10888277720879 (each), 20888277720876 (case); ALL LOTS

Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →