Medline Industries, LP
What this record is not
A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
MEDLINE Urethral Catheter, Red Rubber Latex, Medline Product Number/SKU (REF), French sizes: 1. DYND13508 - 08 Fr; 2. DYND13510 - 10 Fr; 3. DYND13512 - 12 Fr; 4. DYND13514 - 14 Fr; 5. DYND13515 - 15 Fr; 6. DYND13516 - 16 Fr; 7. DYND13518 - 18 Fr; 8. DYND13520 - 20 Fr; 9. DYND13522 - 22 Fr; 10. DYND13614 - 14 Fr; 11. DYND13616 - 16 Fr.
1. DYND13508, UDI/DI 10080196031139 (each), 20080196031136 (case); 2. DYND13510, UDI/DI 10080196726462 (each), 40080196726463 (case); 3. DYND13512, UDI/DI 10080196031177 (each), 40080196031178 (case); 4. DYND13514, UDI/DI 10080196726219 (each), 40080196726210 (case); 5. DYND13515, UDI/DI 10080196031214 (each), 40080196031215 (case); 6. DYND13516, UDI/DI 10080196031221 (each), 40080196031222 (case); 7. DYND13518, UDI/DI 10080196031245 (each), 40080196031246 (case); 8. DYND13520, UDI/DI 10080196031252 (each), 40080196031253 (case); 9. DYND13522, UDI/DI 10080196061259 (each), 40080196061250 (case); 10. DYND13614, UDI/DI 10080196031276 (each), 30080196031270 (case); 11. DYND13616, UDI/DI 10888277720879 (each), 20888277720876 (case); ALL LOTS
Data: U.S. FDA. Recall details are as reported by the FDA. The quality-domain mapping and severity are a SPEQ editorial interpretation of the recall reason, not an official FDA classification. Open the enforcement explorer →