Z-2398-2026Class II

Siemens Healthcare Diagnostics, Inc.

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product type
Device
Status
Ongoing
Report date
2026-06-17
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Brazil, Croatia, Germany, Hungary, India, Italy, Japan, Korea, Lithuania, Pakistan, Poland, Saudi Arabia, Spain, South Africa, Sri Lanka, Turkey, United Arab Emirates.
Firm state
DE
Product quantity
2878 units
Initiated by
Voluntary: Firm initiated
Reason for recall

A potential positive bias in quality control (QC) and patient sample results in three lots of the Dimension Tacrolimus (TAC) Flex reagent cartridge.

Product description

Siemens Dimension Tacrolimus Flex Reagent Cartridge (TAC), REF DF207, SMN 10700795; enzyme immunoassay

Code / lot info

Manufacturing Lot/Serial Numbers: FA6254, FA6260, FA6286; UDIs: (01)00842768035425(10)FA6254(17)20260911; (01)00842768035425(10)FA6260(17)20260917; (01)00842768035425(10)FA6286(17)20261013.