← FDA ENFORCEMENT
Z-2355-2026Class II

In2bones USA, LLC

What this record is not

A recall record is reproduced as FDA published it, and the classification is FDA’s. SPEQ adds no interpretation, and a recall concerns the products and dates in the record — it is not a statement about a company’s other products or its current status.

Recalling firm
In2bones USA, LLC
Product type
Device
Status
Ongoing
Report date
2026-06-17
Distribution
U.S.: AL, AZ, CA, CO, FL, GA, HI, ID, IL, IN, LA, MD, MO, NC, NJ, NV, NY, OH, OK, PA, TN, TX, UT, VA, and WI O.U.S.: Belgium, Canada, Colombia, France, Ireland, Netherlands, Spain, United Kingdom,
Firm state
TN
Product quantity
306 kits
Initiated by
Voluntary: Firm initiated
Reason for recall

Due to the Magnetic Resonance Conditional Symbol missing on the carton, inner tray, and patient label.

Product description

NeoSpan¿ SuperElastic Compression Staple w/Instruments Catalog Number and Product Name T50 SN010 NeoSpan¿ SuperElastic Compression Staple w/Instruments, Size 10x10x10 UDI-DI Code: 00817906020083 T50 SN012 NeoSpan¿ SuperElastic Compression Staple w/Instruments, Size 12x12x12 UDI-DI Code: 00817906020090 T50 SN020 NeoSpan¿ SuperElastic Compression Staple w/Instruments, Size 20x20x20 UDI-DI Code: 00817906020151 T50 SN110 NeoSpan¿ SuperElastic Compression Staple w/Instruments, Size 10x15x13 UDI-DI Code: 00817906020113 T50 SN115 NeoSpan¿ SuperElastic Compression Staple w/Instruments, Size 15Wx15x15 UDI-DI Code: 00817906020137 T50 SN118 NeoSpan¿ SuperElastic Compression Staple w/Instruments, Size 18x14x14 UDI-DI Code: 00810021861364 Single/multiple component metallic bone fixation appliances and accessories The In2Bones USA NeoSpan" Compression Staple Implant w/instruments is indicated for hand and foot bone fragments, osteotomy fixation and joint arthrodesis.

Code / lot info

Catalog No. Lot Code T50 SN010 1962242 1964665 2054541 2039361 T50 SN012 1963511 T50 SN020 1971252 1990642 2027441 T50 SN110 1971241 T50 SN115 19647210 T50 SN118 1990631